|
|
![]() |
|
|
Ebeid MR. Balloon expandable stents for coarctation
of the aorta: review of current status and technical considerations. Images
Paediatr Cardiol 2003;15:25-41
|
| coarctation | stent | catheter intervention |
| heart disease, congenital |
Stents
After the initial success in treating iliac artery stenosis,18-20
stents were used to treat stenotic lesions associated with congenital heart
disease 21-23 both native and post operative and shortly there
after were used in coarctation of the aorta.11-17 The most commonly
used stent is the balloon dilatable type which allows for further expansion
of the stent as the patient grows. Self expanding stents have been only
infrequently used because of the limitation of further expansion and the
limited radial strength.24, 25 The following are the stents
commonly used for coarctation of the aorta.
I-Palmaz stents (Johnson & Johnson Interventional Systems Co.,
Warren, NJ)
The Palmaz stents are balloon expandable stainless steal prostheses.
They have a closed cell design which gives them high radial strength but
makes them less flexible. Most of these stents are hand mounted and crimped
on the delivery balloon. Until recently the large Palmaz stents (P308,
188) were the only ones available in the United States and considered suitable
for coarctation of the aorta. The first 2 digits are the length of the
stent in mms. The final digit represents the minimum recommended expanded
diameter. This in turn reflects on the maximum diameter it can achieve
and the radial strength. The wall thickness is .005 inch and the unexpanded
diameter is 3.4 mm. All these factors influence the size of the balloon
it can be crimped on as well as the minimum sheath size. These stents have
a recommended expanded diameter between 8-12 mm but are commonly over dilated
to 18-20 mm with some difficulty. A significant number of patients have
had P308 stents implanted in the coarctation site with excellent results12-16
(figure 1 and 2, Table I). However over dilation of these stents results
in significant foreshortening < 33% at 12 mm and up to 50% at 18 mm15
and is met with significant stent resistance. Smaller Palmaz stents (series
ending with the digit 4) have been used sporadically under individual circumstances
and are not generally used because of the limited expandability with growth.
Extra large diameter Palmaz stents
These have became available in the United States and were designated
P3110, 3010, 4010, 5010, (4014, 5014 outside the US). The numbering implications
are similar to the 8-series outlined earlier. The wall thickness is .013
in. These stents are expandable to 25 mm while maintaining their radial
strength. Their foreshortening ranges between 2.5% at 10 mm and 23% at
25 mm (off label). The 3010 is available balloon mounted on 12 mm delivery
balloon. The others are hand mounted on the selected balloons. These stents
require relatively larger balloons for adequate hand crimping and larger
sheaths than the series ending with 8. (figure 3-5, tables 1 & 2).
They offer a significant advantage because of their expandability, radial
strength and less fore shortening.
The Palmaz Genesis stent
The Palmaz-Genesis stent (Cordis / Johnson and Johnson) was introduced
to overcome some of the short falls of the Plamaz stents. It was designed
to have lower profile, minimize shortening, add radial strength, and increase
flexibility. It has a closed cell design based on the FlexSegemnt technology
(Figure 6) which improves flexibility, reduces shortening and improves
scaffolding in a bend. The original Genesis Ò
stent was successful in negotiating tight curves and had good radial strength.
However the maximum diameter which can be achieved is less than optimal
for coarctation. Subsequently the Palmaz Genesis, large and extra large
diameters (XD), were introduced to allow larger diameter expansion while
retaining the favorable characteristics of the genesis stent. The Genesis
XD is expandable off label to 18 mm in bench work while maintaining flexibility
and excellent radial force. (Tables 1 and 2). It has equivalent or higher
strength to the Palmaz P-308 stent with less shortening. Its application
and expandability for coarctation remains to be tested in a large group
of patients.
II-The Intrastent (IT) double strut Ò
(Intratherapeutics, St. Paul, MN, recently acquired by ev3, Plymouth,
MN)
The IT LD stent was introduced in an attempt to improve on the only
stent existing at that time considered suitable for coarctation (namely
the Palmaz). This new stent has an open cell design composed of parallel
struts of laser cut 316L stainless steel.26, 27 The unexpanded
diameter is 3.8 mm length with no foreshortening at 12 mm diameter. The
design allows for better flexibility and maneuverability. The maneuverability
may be a significant advantage when placing the stent in tortuous areas
through tight curves as the pulmonary arteries, though this is not necessarily
the case in coarctation. The flexibility does, however, offer some possible
advantage when implanted along the curve of the aortic arch. The initially
introduced Intrastent (double strut LD) was tested in bench research up
to 12 mms. against the Palmaz stent, and was felt to have comparable radial
strength. Expanding the stent beyond 12 or 15 mms. has not been as encouraging
as what initially was hoped. It was noted to have less radial strength,
26, 27 (and personal experience). Elastic recoil of the stent
was noticed on deflation of the balloon in certain lesions requiring high
radial strength. Additionally, distortion of the stent with either shortening
or elongation has been recognized when larger diameters were attempted.
This stent has not gained popularity in coarctation which may require diameter
beyond 12-15 mms. The same company introduced two stents which are expandable
to larger diameters while maintaining the integrity of the stent as well
as the radial strength. The Mega LD Ò
(unexpanded diameter 3.8 mm) and Max LD Ò
(unexpanded diameter 4.5 mm) stents were introduced (series S17 and S18
respectively ) in varying lengths (16, 26, 36 mm). These stents can be
expanded up to 18 and 25 mms., respectively, with < 25 % shortening
at the largest diameters. They also maintain good radial strength. These
stents have the possible advantage of conforming to the curve of the aortic
arch, (figures 7-9) while maintaining the radial strength as well as the
advantage of open cell design especially if crossing a side branch. These
however require slightly larger balloons and sheaths for adequate crimping
and deployment because of their larger unexpanded diameter compared to
the Palmaz stents (Series-8).
III -Cheatham-Platinum (CP) Stent (NuMed, Hopkinton, NY)
The CP stent was developed by NuMed (Hopkinton, New York), in collaboration
with John Cheatham, MD, to provide improvement on the stents available
at that time which was the Palmaz. This stent was made from heat-tempered
90% platinum and 10% Iridium, .01 inch wire arranged in a zig pattern.16
The number of the zigs is variable between 6-8 per row and will affect
the final diameter, profile and the strength of the stent as well as the
percent shortening. If the target lesion expected diameter is < 15 mm
then a 6 zig would be sufficient especially if it is in small child thus
having a lower profile stent. The latter, however, would have limited application
for coarctation because of the usually larger diameter required. The 8
zig / row configuration can be dilated up to 24 mm but has larger profile
(28) and would offer a better choice for treating coarctation. A custom
made 10 zig stent can be made and dilatable up to 30 mm with < 20 %
shortening but would have a larger profile requiring larger sheath. Initial
published results of this stent are encouraging. 16 There was
rare incidence of stent fatigue fracture.
Technical tips for implantation:
This section is not intended to be a full discussion of the implantation
technique, but rather to present a few tips reflecting personal preference
and may or may not be applicable to every patient or operator. We, however,
feel that these are helpful to achieve successful stent implantation.
1. Wire choice
The wire is chosen to provide extra support (for example: extra stiff
or super stiff wires) to avoid stent / balloon migration during implantation.
We position the wire to ensure that the stiff part is well in the position
of implantation and extends proximally as much as possible keeping the
floppy end distant from the
implantation site. This is accomplished by placing the tip of the wire
in the right subclavian artery (or in the left subclavian, depending on
the location of the coarctation) or occasionally by making a loop in the
ascending aorta. (figures 2, 4, 5, 7, 8)
2. Complex arch anatomy
In these patients, the position of the stent has to be very accurate
to avoid unnecessary covering of the proximal branches or extension to
distal aneurysms. We usually enter the left ventricle from prograde fashion,
either using a transseptal puncture (figure 2, 4, 5, 7) or through a PFO
(figure 8) and place an angiographic catheter in the left ventricle to
obtain angiograms and pressure measurements, before, during and after deployment.
Multiple angiograms are obtained; before and after uncovering the stent
with adjustment of stent position as necessary.
3. Stent choice
The ideal stent is probably not available, but certain characteristics
make some stents more attractive than others in different positions of
the coarctation. The stent should be chosen that is expandable to an adult
size aorta. A more flexible stent may be better suited if the implantation
is across the arch ( figures 7, 8).
If the stent may cover side branches or neck vessels, then an open
cell design may be more attractive.
4. Balloon choice and inflation
The perfect balloon would be of low profile, high pressure, have enough
material to allow easy crimping of the stent and resistant to rupture.
These facts are difficult to achieve because a thin balloon would be more
prone to rupture, and more difficult to accurately mount and crimp the
stent. A larger material balloon will require a larger sheath. However
the industry has achieved significant improvement in the balloon designs
and material. Another consideration is the length of the balloon in relation
to the stent i.e. shorter versus longer. Both offer some potential advantages
and is interventionalist dependent. The shorter balloons offer the advantage
of inflating the inner part of the stent first thus avoiding the potential
complications of flaring of the ends of the stents resulting in balloon
rupture or vessel wall perforation. However it can result in stent sliding
or embolization making it difficult to accurately place the stent. The
longer balloons allow more precise placement and repositioning if necessary
before fully inflating the balloon. We generally use balloons slightly
longer than the stent or the BIB balloon (NuMed, Hopkinton, NY), which
offers two balloons, one shorter (inner) and one longer (outer) than the
stent. When we use the longer balloon method and if precise positioning
of the stent is required, we find it helpful to initially uncover only
the distal part of the balloon (the end further from the operator). The
balloon is slightly inflated, thus allowing the distal part to slightly
expand (Figure 8C). We obtain further pictures and reposition, if necessary.
When we are comfortable with the position, the balloon and stent are fully
uncovered and further mild inflation is done to allow the proximal part
of the balloon (the end close to the operator) to inflate (figure 8D).
This is done before fully inflating the balloon, thus the stent is held
in accurate position by the two partially inflated ends of the balloon.
An angiogram is obtained to confirm the position of the stent and readjustment
is done as necessary with repeat angiograms. When the position is accurate,
the stent is fully inflated. During the inflation, steady pressure is maintained
by the operator on the balloon to ensure the balloon/stent will not travel.
Maintaining the extra support wire in that position, as well as the steady
hand of the operator with gentle push, should maintain the stent in its
position. If during the inflation, there is any movement of the balloon,
this can also be readjusted by minimal hand manipulation or alternatively
to stop the inflation and reposition before the full inflation. Alternatively,
the BIB balloon allows inflating the center of the stent and the inner
balloon first (Figure 4). This also may allow partial repositioning of
the stent before the full inflation. Once the stent has been deployed,
repeat angiograms are obtained from the catheter in the LV to confirm there
is no dissection and the accurate position of the stent. The pressure from
the LV is compared with the pressure obtained from the arterial end thus
estimating the residual gradient. Based on the result of the angiogram
and the residual gradient, if any, mild further dilatation can be done.
Achieving a perfect result on the first time implanting the stent especially
if this is a primary stenting is not necessary. Further dilatation after
six months to a year can be done.
5. Extent of dilatation
When implanting the stent, the goal is to achieve the largest possible
diameter and eliminate any gradient to decrease the need for redilation.
However the concerns are always the possibility of tear or aneurysm formation
which may be encountered with over dilation We have adopted certain criteria
to help achieve safe and effective stent implantation. We measure carefully
the descending aorta at the level of the diaphragm and the proximal transverse
arch. The balloon is genrerally chosen to be equal to the lesser of the
2 measurements not to exceed 3 times the size of the stenotic segment.
The balloon is inflated to expand and deploy the stent which may not be
the full diameter of the balloon or the full pressure. When the stent is
expanded well we repeat angiography and pressure measurement. If there
is significant residual gradient and no dissection, repeat balloon dilation
using the same balloon at the maximum pressure is done. Repeat assessment
is performed. If there is still significant gradient and no dissection,
and we have not exceeded 3 times the size of the of the stenotic segment,
a slightly larger balloon may be used. Subsequently flaring of the ends
using a lower pressure inflation may be performed if necessary though it
is not essential. Repeat assessment is done. If there is significant residual
gradient we would be reluctant to perform further dilation at this point.
The patient would be brought back to the catheterization laboratory for
repeat dilation at a later date. We like to wait at least 6 months. Achieving
a perfect result is not necessary at the time of stent implantation.
6. Anticoagulation
Patients are maintained on aspirin for at least one year after the
procedure to avoid platelet aggregation. Since this is a high pressure,
high flow area it was not felt necessary to use warfarin to anti-coagulate
the patients.14
Discussion
Coarctation, if left untreated, poses a significant health risk and
up to 90% of the patients with isolated coarctation may die before age
fifty years of complications of coarctation including aortic rupture, intracranial
hemorrhages, hypertension, endocarditis or heart failure. 29, 30
With treatment of coarctation and aggressive management of this disease,
significant improvement in the long-term survival of these patients has
been seen and is hopefully going to improve with effective and successful
management. The major burden on the interventional cardiologist is to choose
the right approach for the specific patient, whether surgery, balloon or
stent. Each of these modalities have a role for the specific patient. The
patient's age weight history of surgery influences in great deal the management
strategy. Though the use of stents to treat coarctation of the aorta has
been relatively recent, its use to treat peripheral vascular stenosis dates
to the mid 1980s 18-20 and later in the management of other
stenotic lesions in congenital heart disease both native and post operative.21-23
Early animal work in experimental coarctation model 21,32 showed
promising results. However, in another animal model few instances of rupture
occurred with redilation.33 It appeared that in this specific
animal model the rupture occurred when the dilation was attempted at significantly
higher balloon diameter to stenotic lesion ratio. Slowly but steadily the
use of stents was introduced in the management of coarctation. The initial
and follow up results have been encouraging with successful gradient reduction.
12-17, 34 Similar to redilation in other lesions, 32
redilation of the stented coarctation has been possible with further gradient
reduction. Few complications however have been observed.11, 15, 16,
36. Some appear to be related to using a significantly larger balloon
to stenotic site ratio or related to balloon choice. Recoarctation following
balloon angioplasty alone has been seen in 18-31 % of patients with native
coarctation and in 9-80% of postoperative coarctation, 8,9,11
depending on the age of the patient, the nature of the lesion and size
of balloon used. Late follow up of angioplasty patients showed that 32%
of the patients with recurrent coarctation had not had adequate gradient
relief (<20 mm. Hg) at the initial angioplasty despite adequate balloon
/ isthmus ratio.8, 11 Intravascular stents can maintain the
patency of a stenotic lesion by eliminating the elastic recoil of the tissue
and can extend over a relatively long segment of stenosis, thus decreasing
the need for over expansion or causing intimal tear in these vessels. 10,
37 Short and intermediate term follow up have shown that stented
coarctation have good ( if not better) gradient relief 12-15
and usually the recurrence is attributed to somatic growth 14, 17
or conservative initial dilation (personal experience) to avoid using too
high balloon to stenosed segment ratio exceeding (3-4 times) in the initial
implantation. In few patients 17 neointimal growth causing mild
restenosis was seen in growing children who were stented < 6 years old.
Extension of a stent into the orifices of the branches of the aortic arch is a concern as the struts can cause obstruction to the flow in major vessels. In adult size patients, however, the struts of a dilated stent are very small relative to the diameter of the subclavian and carotid arteries. Published reports for coarctation as well as in other sites 14,36 have indicated continued patency and unimpeded flow in these vessels. This however must be taken into account during stent implantation. Every effort should be made to minimize the extension of the stent across the side branches if possible. In those instances where a stent will have to cross the side branches there may be a potential advantage for the open cell design though the experience with these stents in that position is limited at this time.
In conclusion, in selected patients with coarctation of the aorta stent
implantation may be a feasible and improved option to relieve the stenosis.
Short and mid term follow-up of these patients have shown encouraging results.
Further studies, especially addressing long term follow up, are needed.
References
|
|
|
|
|
![]() |